It Shouldn't Work: Imaging Actinium-225 on a PET Scanner

Dustin Osborne • October 16, 2026

It Shouldn't Work: Imaging Actinium-225 on a PET Scanner

Some of the best physics starts with a rule you were taught and a colleague willing to test it.

Every nuclear medicine physicist, technologist, and resident I have taught learned the same rule early: PET images positron emitters. A positron annihilates, two 511 keV photons leave back to back, the scanner draws a line between the two detectors that caught them, and the image builds up from millions of those lines. Actinium-225 is an alpha emitter. No positrons, no 511 keV pairs, no lines. If you had asked me five years ago whether you could image Ac-225 on a clinical PET scanner, I would have handed you that rule as the answer.

A group of colleagues just published the paper showing why that answer was incomplete.

What the team did

In a paper published September 30 in the Journal of Nuclear Medicine , Javier Caravaca (Lawrence Berkeley National Laboratory and UCSF), Youngho Seo, Robert Flavell, and Thomas Hope (UCSF), and Jorge Cabello and Stefan Siegel (Siemens Medical Solutions, here in Knoxville) imaged Ac-225 on a Biograph Vision 600 PET/CT running in research mode. They worked through vial and clinical-size phantom experiments, compared the results against SPECT, and then imaged a patient with prostate cancer after treatment with [ 225 Ac]Ac-PSMA-617. That is the first image of an Ac-225 therapy agent in a patient acquired on a PET scanner. They call the method time-of-flight cascade gamma-ray imaging, or TOF-CGI, and Berkeley Lab received a 2026 R&D 100 Award for it in August.

I will admit some local bias. I spent years at Siemens, and the scanner this team pushed well past its intended use was developed here in Knoxville. It is a real pleasure to watch people you know take a machine you know well and make it do something none of us expected.

The physics, briefly

The trick lives one branch down the Ac-225 decay chain. About 2% of Bi-213 decays go by alpha emission to Tl-209, and Tl-209 beta decays to Pb-209 through a fast gamma cascade: a 465 keV photon followed by a 1567 keV photon, with the intermediate nuclear state having a half-life of about 45 ps. For practical purposes the two photons leave the decay site at the same instant.

They do not leave back to back. The two directions are uncorrelated, so there is no line of response. What remains is timing. If two photons start at the same moment from the same point, the difference in their arrival times fixes the difference in the distances they traveled. The set of points satisfying that condition is a hyperboloid with the two detectors at its foci, and the scanner's coincidence timing resolution sets how thick that surface is.

This is the part I find most satisfying. It is the same equation TOF-PET already solves. Conventional time-of-flight uses the arrival time difference to place an annihilation somewhere along its line, and the line comes for free from collinearity. Take collinearity away and the timing still constrains the decay to a surface. The reconstruction has to work much harder, which is why Caravaca wrote the algorithm from scratch and used GPU computing at NERSC to make it practical, but the physical information was sitting in the data stream all along.

Timing performance was the enabling condition. The team measured 205 ps FWHM coincidence timing for the Tl-209 cascade, slightly better than the 215 ps measured with annihilation photons on the same system. At that resolution the surface is about 3 cm thick. With the 500 to 600 ps timing of the first generation of commercial TOF scanners, it would have been two to three times thicker.

Line or surface?

A 2D slice through an 80 cm detector ring. Tap inside the ring to move the source. Each event shades where the timing says the decay could have been. Stack events and watch the image form.

Timing resolution (FWHM) 205 ps
Events shown 1

Simplified 2D illustration: no attenuation, scatter, detector size, or depth of interaction. The 205 ps default is the cascade timing measured by Caravaca et al.

What the numbers say

In phantoms, TOF-CGI detected Ac-225 with an efficiency of 0.13%, compared with 0.016% for the SPECT acquisition, roughly an eightfold difference. Signal-to-noise ratio was higher than SPECT at every cylinder size and scan time tested, and TOF-CGI reached that higher SNR with a scan time ten times shorter. The minimum imageable activity was 5.5 kBq in a 20-minute scan, and the response was linear with activity (r 2 = 0.997). In the patient, TOF-CGI found the same lesions seen on the pretherapy [ 68 Ga]Ga-PSMA-11 PET and produced a higher SNR image than SPECT.

One more point deserves attention. Because the signal comes from Tl-209, the image shows where the decay chain actually is after any daughter release and redistribution. A positron-emitting surrogate only tells you where the targeting molecule went. Daughter redistribution is one of the central open questions in Ac-225 dosimetry, and this is a direct look at it.

What it does not do yet

The authors are clear about the limits, and they are real. This is one patient. The reconstruction does not yet include attenuation, scatter, or sensitivity corrections, so the images are not quantitative, and recovery coefficients came in lower than SPECT largely for that reason. Spatial resolution was about 2 cm transaxially and 4.2 cm axially, with the axial number limited by the scanner's short axial field of view. The scanner ran in a research configuration, and the authors note that clinical use would need more evidence and regulatory work. They point to long axial field-of-view systems and better timing as the obvious routes to improvement, and the team has named Pb-212 and Lu-177 as other cascade emitters worth testing.

None of that diminishes the result. A feasibility paper is supposed to show that the door exists. Walking through it is the next decade of work.

"Shouldn't work" is a hypothesis

A conversation I had with a student recently kept coming back to me while I read this paper. We were talking about how much of what we teach is a set of constructs: models that hold within the conditions where we learned them. They are necessary. You cannot train a physicist, a technologist, or a physician without them. The risk is that a model learned well enough starts to feel like a law, and its boundary conditions quietly disappear.

"PET needs positrons" is a good example. Stated carefully, it describes how conventional reconstruction uses collinear 511 keV pairs. It was never a statement about what a ring of fast scintillators with sub-nanosecond timing can physically detect. Someone had to ask which assumption the rule actually rested on and then spend the time finding out. By Caravaca's own account to Berkeley Lab, the project started with the team wondering whether they could put actinium's gamma-ray pairs to work in a PET scanner.

My takeaway from that conversation, and from this paper, is to treat "that shouldn't work" as a hypothesis. Most of the time the experiment confirms the textbook, and you come away understanding the textbook better than you did. Occasionally it does not, and that is where new things come from. Science moves forward when people are willing to spend some of their time in the category of ideas that probably will not pan out.

Why this matters for community programs

Most of ARC's work is operational: dosimetry, staffing, regulatory compliance, referral pathways. Ac-225 agents are moving through clinical trials now, and imaging them well is an unsolved problem at academic centers, let alone community sites. I do not expect anyone in the corridor to be running TOF-CGI next year. The longer lesson is that the equipment already installed in our departments may be capable of more than its original specification suggests, and the people who run it every day are often the ones best positioned to notice. Some of the most practical questions in theranostics will be answered by technologists and physicists who ask why a rule exists before accepting that it does.

What rule from your own training have you watched turn out to be a boundary condition instead of a law?

Sources

Caravaca J, Seo Y, Cabello J, Siegel SB, Flavell RR, Hope TA. Clinical Imaging of 225 Ac in a PET Scanner via Time-of-Flight Cascade γ-Ray Imaging. J Nucl Med. Published online September 30, 2026. doi:10.2967/jnumed.126.272594

Lawrence Berkeley National Laboratory. Targeted Cancer Therapy Gets a Sharper Focus. October 1, 2026.

Lawrence Berkeley National Laboratory. Berkeley Lab Technology Wins R&D 100 Award. August 11, 2026.

By Dustin Osborne • November 26, 2026
By Dustin Osborne • November 13, 2026
mountains
You’ve chosen a terrific way of integrating images and text into your website. Move the image anywhere you want in this container and the text will automatically wrap around it. You can display events team members new products and more easily and creatively. To start add an image from the Image Picker and edit it as you would edit any image in the system. For example you can link the image to existing pages in your site a website URL a popup or an anchor. After you’ve chosen the image add your text. You can add text that describes the image you’ve selected or simply use the image for decorative purposes. \nYou’ve chosen a terrific way of integrating images and text into your website. Move the image anywhere you want in this container and the text will automatically wrap around it. You can display events team members new products more easily and creatively. To start add an image from the Image Picker and edit it as you would edit any image in the system. For example you can link the image to existing pages in your site a website URL a popup or an anchor. After you’ve chosen the image add your text. You can add text that describes the image you’ve selected or simply use the image for decorative purposes.